T9DCS/T10DCS Double Monochromator UV-Vis
T9DCS: UV-Vis Spectrometer
T10DCS: UV-Vis Spectrometer
The T9/T10DCS Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy.
The optical design of both the T9DCS and T10DCS offer extremely low stray light characteristics (≤0.00004%T NaI,220 nm) which allows for an extensive photometric range (-8.0 – 8.0Abs). Measurements at deep ultra-violet wavelengths can also be achieved with use of Nitrogen purged optics.
The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel.
Key Features
◆ True Double Beam Double monochromator Optics, and fully sealed to provide ultra-low stray light (0.00004%%T @220nm) and high photometric range (-8.0~8.0 Abs)
◆ 185– 900nm Wavelength range with Nitrogen purge
◆ Precise Wavelength accuracy is ensured by the integrated Mercury lamp for automatic correction
◆ Continuously selectable spectra bandwidth from 0.1-5nm
◆ Beam size continuously adjustable for variety of applications including ultra-volume cuvettes
◆ Large sample compartment suitable for all kinds of accessories, especially for 150mm integrating sphere
◆ Support WIFI communication in field operation
◆ The integrated Mercury Emissions Lamp with automatic correction function ensures precise wavelength accuracy
T9/T10DCS Double Monochromator UV-Vis Spectrometer - Brochure
Specifications
Model | T9DCS | T10DCS |
Optical System →Click here to know more about | Double beam | Double beam |
Scan Speed | 1000 nm/min 1500 nm/min location speed | 1000 nm/min 1500 nm/min location speed |
Wavelength Range | 185-900 nm | 185-900 nm |
Wavelength Accuracy | ±0.2 nm | ±0.2 nm |
Spectral Bandwidth | ±20 % (0.1 nm-5 nm continuously adjustable) | ±20 % (0.1 nm-5 nm continuously adjustable) |
Photometric Mode | Transmittance,Absorbance, Enegy and Concentration | Transmittance,Absorbance, Enegy and Concentration |
Photometric Range | -6.0 Abs-6.0 Abs | -8.0 Abs-8.0 Abs |
Photometric Accuracy | ±0.004 A@2.0 A ±0.003 A@1.0 A ±0.002 A@0.5 A ±0.3 % | ±0.004 A@2.0 A ±0.003 A@1.0 A ±0.002 A@0.5 A ±0.3 % |
Photometric Reproducibility | ≤0.002 A@2.0 A ≤0.0008 A/1.0 A ≤0.0004 A/0.5 A ≤0.1 % | ≤0.002 A@2.0 A ≤0.0008 A/1.0 A ≤0.0004 A/0.5 A ≤0.1 % |
Photometric Noise | 0 % Noise: ≤0.01 % 100 % Noise: ≤0.1 % | 0 % Noise: ≤0.01 % 100 % Noise: ≤0.1 % |
Baseline Flatness | ±0.0008 Abs | ±0.0005 Abs |
Baseline Stability | 0.1 %/hr | 0.1 %/hr |
Stray light | ≤0.0001 %T (NaI, 220 nm) ≤0.0001 %T (NaNO2, 360 nm) | ≤0.00004 %T (NaI, 220 nm) ≤0.00002 %T (NaNO2, 360 nm) |
Cell Holder | Automatic 8 Cell Changer | Automatic 8 Cell Changer |
Detector | Photo multiplier tube | Photo multiplier tube |
Light Source | D2 lamp,Tungstern lamp,Mercury lamp for WL correction | D2 lamp,Tungstern lamp,Mercury lamp for WL correction |
PC Interface | RS232, USB, Wifi | RS232, USB, Wifi |
Software Support | UV Win | UV Win |
Power Supply | Switchable 120-230 VAC 50-60 Hz | Switchable 120-230 VAC 50-60 Hz |
Weight | 136 lb | 136 lb |
Dimensions (Width, Depth, Height) | 36 x 25 x 12 in | 36 x 25 x 12 in |
Accessories & Spare Parts
-
UVWIN Software
Key Functions:
-Spectral Analysis
-Quantitative Analysis
-Kinetic Analysis
-DNA Protein Determination Analysis
-
Sipper Pump Accessory
Description:
Double channels sipper pump achieves on-line sample analysis. It’s very convenient to perform on-line analysis combined with our UVWin software.
-
Constant Temperature Fixed-cell Holder
Description:
Constant temperature fixed-cell holder is used together with constant-temperature water bath and water pump.
They can measure samples at a specific temperature. And we can provide optional Peltier Accessory for constant temperature circle water.
-
Dissolution Accessory
Description:
Double beam dissolution accessory is used together with double beam automatic 8-cell holder for dissolution rate measurement of sample.
-
100mm Long Pathlength Holder
Description:
The 100mm double beam long path length cell holder can be placed with 5-100mm cuvettes. It is easy for the measurement of samples with long pathlength.
-
Automatic 8-cell Holder
Description:
The automatic 8-cell holder is designed for multi-sample measurement. 8 samples can be put in the cell holder for separate measurements.
-
Test Tube Holder
Description:
Double beam test tube holder is used to hold test tubes. It is easy for measuring of liquid sample in test tubes.
-
Micro Cell Holder
Description:
This is a pair of adjustable base for micro cuvette in the double beam micro cell holder.
With micro cuvette, it can improve measurement precision for micro volume sample.
-
50mm Long Pathlength 5-cell Holder
Description:
Double beam 50mm long pathlength 5-cell holder is easy for multi-sample measurement at one time. Max. Five long pathlength cuvettes can be put in the cell holder for respective measurement. The pathlength of the cuvette can be 5-50mm. It is easy for the measurement of samples with long pathlength.
-
Specular Reflection Accessory
Description:
The specular reflection accessory can measure the reflectance spectrum of solid surface, such as optical surface, crystal surface, and film-coated lens and so on.
-
Angle-changeable Solid Sample Holder
Description:
Slice solid sample can be put on the angle-changeable solid sample holder and fixed on the optical circuit. The angles can be adjusted within±45°. It is easy to measure the absorbency for solid samples from multi-angles.
-
Solid Sample Holder
Description:
Slice solid sample can be put on the double beam solid sample holder and fixed on the optical path. It is easy to measure the absorbance spectrum for solid samples.
-
Double beam 60mm & 150mm integration sphere
Description:
Incident angle: Sample 0°,reference 8°
Minimum dimensions: 15mm×25mm Diffuse reflectance
Minimum dimensions: Diameter 20mm Transmittance
Wavelength range: 230-850nm 5nm slit
Sphere diameter: 150mm