HOME PRODUCTS UV/VIS SPECTROPHOTOMETER T9DCS/T10DCS Double Monochromator UV-Vis

T9DCS/T10DCS Double Monochromator UV-Vis

T9DCS:      UV-Vis Spectrometer        

T10DCS:    UV-Vis Spectrometer        




The T9/T10DCS Spectrophotometer incorporates dual monochromator technology making it well suited to even the most demanding of applications in all areas of UV-Visible Spectroscopy.

 

The optical design of both the T9DCS and T10DCS offer extremely low stray light characteristics (≤0.00004%T NaI,220 nm) which allows for an extensive photometric range (-8.0 – 8.0Abs). Measurements at deep ultra-violet wavelengths can also be achieved with use of Nitrogen purged optics.

 

The instrument can be optically configured to suit the needs of the sample by means of a continually adjustable slit for precise control of spectral resolution and beam size adjustment by means of an attenuating wheel. 



Key Features



◆ True Double Beam Double monochromator Optics, and fully sealed to provide ultra-low stray light (0.00004%%T @220nm) and high photometric range (-8.0~8.0 Abs)

 

◆ 185– 900nm Wavelength range with Nitrogen purge

 

◆ Precise Wavelength accuracy is ensured by the integrated Mercury lamp for automatic correction

 

◆ Continuously selectable spectra bandwidth from 0.1-5nm


◆ Beam size continuously adjustable for variety of applications including ultra-volume cuvettes

 

◆ Large sample compartment suitable for all kinds of accessories, especially for 150mm integrating sphere

 

◆ Support WIFI communication in field operation


◆ The integrated Mercury Emissions Lamp with automatic correction function ensures precise wavelength accuracy


T9/T10DCS Double Monochromator UV-Vis Spectrometer - Brochure


The Comparison Data Sheet  T9DCS/T10DCS UV and Competitors

Specifications

ModelT9DCST10DCS

Optical System

→Click here to know more about 

Optical Path System←


Double beamDouble beam
Scan Speed1000 nm/min
    1500 nm/min location speed
1000 nm/min
    1500 nm/min location speed
Wavelength Range185-900 nm185-900 nm
Wavelength Accuracy±0.2 nm±0.2 nm
Spectral Bandwidth

→Click here to know more about 

SBW←

±20 % (0.1 nm-5 nm continuously adjustable)

±20 % (0.1 nm-5 nm continuously adjustable)
Photometric Mode

Transmittance,Absorbance,

Enegy and Concentration

Transmittance,Absorbance,

Enegy and Concentration

Photometric Range -6.0 Abs-6.0 Abs -8.0 Abs-8.0 Abs
Photometric Accuracy ±0.004 A@2.0 A

 ±0.003 A@1.0 A

 ±0.002 A@0.5 A

    ±0.3 %
±0.004 A@2.0 A
±0.003 A@1.0 A
±0.002 A@0.5 A
    ±0.3 %
Photometric Reproducibility    ≤0.002 A@2.0 A
≤0.0008 A/1.0 A
≤0.0004 A/0.5 A
  ≤0.1 %
        ≤0.002 A@2.0 A
    ≤0.0008 A/1.0 A
    ≤0.0004 A/0.5 A
    ≤0.1 %
Photometric Noise0 % Noise: ≤0.01 %
    100 % Noise: ≤0.1 %
0 % Noise: ≤0.01 %
    100 % Noise: ≤0.1 %
Baseline Flatness±0.0008 Abs±0.0005 Abs
Baseline Stability0.1 %/hr0.1 %/hr
Stray light

→Click here to know more about 

Stray Light←

≤0.0001 %T (NaI, 220 nm)
      ≤0.0001 %T (NaNO2, 360 nm)
≤0.00004 %T (NaI, 220 nm)
    ≤0.00002 %T (NaNO2, 360 nm)
Cell HolderAutomatic 8 Cell ChangerAutomatic 8 Cell Changer
DetectorPhoto multiplier tubePhoto multiplier tube
Light SourceD2 lamp,Tungstern lamp,Mercury lamp for WL correctionD2 lamp,Tungstern lamp,Mercury lamp for WL correction
PC InterfaceRS232, USB, WifiRS232, USB, Wifi
Software SupportUV WinUV Win
Power SupplySwitchable 120-230 VAC 50-60 HzSwitchable 120-230 VAC 50-60 Hz
Weight136 lb136 lb
Dimensions
    (Width, Depth, Height)
36 x 25 x 12 in36 x 25 x 12 in


Accessories & Spare Parts

Related

Other Products

Quote Technical Inquiry